| Sign In | Join Free | My insurersguide.com |
|
Brand Name : Fukuda
Model Number : PCB-6380D-C1
MOQ : 1
Payment Terms : T/T(wire transfer),Western Union,Paypal
Supply Ability : 10
Delivery Time : 5-8 work days
BRAND : Fukuda
PRODUCT NAME : Mainboard
WARRANTY : 90 Days
CONDITION : In Good Working Order
MATERIAL : Plastic, Metal,Wire
COLOR : Green
MOQ : One Piece
IN STOCK : Yes
MODEL : PCB-6380D-C1
PAYMENT TERMS : T/T(wire transfer), Western Union, Paypal
ADVANTAGE : Original
Use for : CardioMax FX-3010 ECG Machine
FUNCTIONALITY : Good
APPLICATION : ECG part
Fukuda CardioMax FX-3010 ECG Machine Mainboard PCB-6380D-C1 Used in Good Working
►Brief Description
The Mainboard PCB-6380D-C1 is the core component of the FX-3010 ECG Machine. This mainboard receives microvolt-level bioelectrical signals from the patient's leads. It coordinates and manages all operations of the machine, including keyboard input, LCD display, thermal printhead drive, motor control, and external communication. It is responsible for analog-to-digital conversion, filtering, analysis, and temporary storage of signals, providing the final waveform data for printing and display. Any ECG signal acquisition-processing-output process must pass through the precise processing of this mainboard.
►Specification
| Brand name | Fukuda |
| Model Number | FX-3010 |
| Product Name | Mainboard |
| MOQ | 1 Piece |
| Condition | Used original |
| Warranty | 90 Days |
| Delivery Method | DHL ,Fedex ,UPS |
| Payment Terms | Full Before Shipment |
| Features | Professional, reliable,Economic |
| Stock Situation | Stock available |
| Packing | Carton |
| Delivery | 3-5 Working Days |
| Method of Transport | Expressage |
| Use for | CardioMax FX-3010 ECG |


►Main Functions of the Motherboard
1. Preamplifier and Lead Selector
Function: This is the first stage of signal processing. It uses a multiplexer to switch different electrode combinations based on the user-selected lead mode (I, II, III, aVR, aVL, aVF, V1-V6). Subsequently, a high-input-impedance, low-noise instrumentation amplifier provides initial amplification of the weak millivolt-level ECG signal.
2. Analog Signal Conditioning Circuit
Function: Further processes the initially amplified signal. This mainly includes:
Filtering: Removing baseline drift (such as that caused by breathing or exercise) using a high-pass filter, and suppressing high-frequency noise such as EMG interference using a low-pass filter. Typically, a powerful power frequency notch filter is also included to specifically filter out 50/60Hz power supply interference.
Amplification: The main amplifier amplifies the signal to a level suitable for processing by the analog-to-digital converter (ADC).
3. Analog-to-Digital Converter
Function: Converting the conditioned, clean analog ECG signal into a digital signal for subsequent digital processing and analysis by the microprocessor. This is a crucial step in the transition of signals from the "physical world" to the "digital world."
4. Processor and Digital Signal Processor
Function: This is the "brain" of the motherboard. It is responsible for:
Executing digital filtering algorithms to further optimize the waveform.
Running ECG analysis programs to perform automatic measurements (such as heart rate and intervals) and arrhythmia analysis.
Thank you for your attention, if you have any questions about Fukuda CardioMax FX-3010 ECG Machine Mainboard PCB-6380D-C1 Used in Good Working, please contact me freely.
|
|